Digital Sight 1000 Ver2.3.6_1.2.8

Digital Sight Series
Software Download Graphic

Software Details

Installation Procedure
This program applies to units with serial numbers 330001 and above. Please make sure to check the serial number before installation.

To install this program, you will need a Digital Sight 1000, an SD card, a monitor, and a mouse.
– Copy the “update.bin” file from the “DS1000_FWv2.3.6_1.2.8” folder to the SD card.
– Insert the SD card into the Digital Sight 1000 and start it in standalone mode.
– A confirmation message for installation will appear. Click “YES” to begin the installation.

Modifications
– Fixed an issue where still images could not be captured when pressing the capture button in NIS-Elements F / L in PC connection mode.
– Fixed an issue where unknown calibration data existed in the factory default state.
– Fixed an issue where the loupe screen was captured when taking still images with the loupe function enabled in standalone mode.
– Fixed an issue where videos could be recorded even when the measurement function was enabled in standalone mode.
– Fixed an issue where CSV and text files containing only numerical measurement results were always saved in English, regardless of the selected language in standalone mode.
– Fixed an issue where the exposure time could not be incremented at certain settings in standalone mode.

Notes
– Units with serial numbers 229999 and below do not experience these issues, so no update is necessary.
– The character encoding for CSV files containing only numerical measurement results is UTF-8. When saving in Japanese, please import the CSV file into Excel as UTF-8.

Release date: 27/5/2025